| Product Specifications |
| Signal Conditioning |
Amplified |
| Flow/Pressure Range |
0 mbar to 5.0 mbar [0 in H2O to 2.0 in H2O] |
| Output Voltage @ Trim Point |
20.0 mA dc ± 1.0 mA dc @ 2.0 in H2O |
| Port Style |
Straight |
| Series Name |
AWM3000 Series |
| Null Shift over Temperature |
± 2.0 mA dc max. |
| Output Shift over Temperature |
- 31% Reading to 24% Reading |
| Maximum change in flow rate |
5.0 SLPM/s |
| Max. Repeatability & Hysteresis
Error |
± 0.50% Reading |
| Null Offset |
4.00 mA dc ± 0.3 mA dc |
| Response Time |
60 ms max. |
| Supply Voltage |
10.0 Vdc ± 0.01 Vdc |
| Maximum Common Mode Pressure |
25.0 psi |
| Power Consumption |
50 mW typ. |
| Operating Temperature Range |
-25 °C to 85 °C [-13 °F to 185 °F]
|
| Storage Temperature Range |
-40 °C to 90 °C [-40 °F to 194 °F]
|
| Media Compatibility |
Dry gas only |
| Weight |
10.8 g |
| Shock |
100 g peak (5 drops, 6 axes) |
| Availability |
Global |
| Comment |
A 5 micron filter must be used on differential
pressure sensors. Temperature shifts when sensing differential
pressure correlates to the density change of the gas over
temperature. Nitrogen calibration gas. |
| UNSPSC Code |
411121 |
| UNSPSC Commodity |
411121 Transducers
|